【Performance evaluation】Cyclic corrosion Test Chamber

Cyclic corrosion Test Chamber

 

Name: Cyclic Corrosion Test Chamber

Model: ZQFS-1200

Manufacturer: Xi'an Tongsheng Apparatus Manufacture Co., Ltd

Purchasing date: December, 2011

Main Specifications:

(1) Temperature range: + 5 ℃ ~ 80 ℃;

(2) Inside humidity range: 40 to 95%;

(3) Sample surface temperature range of the baking zone: + 50 ~ 70 ℃;

(4) Adjustable range of soaking time: 1 to 999 hours;

(5) Adjustable range of baking time: 1 to 999 hours

Common accessories: None

Main application areas: metallic materials

Main functions: experimental research on the corrosion resistance of the steel material

Person in charge of the equipment: Wu Hongyan

Ways of contact:

Tel: 024-83690365

E-mail:

Location: Factory of RAL

 

 

Operation Rules of ZQFS-1200 Cyclic Corrosion Test Chamber

 

  1. Sample loading and solution adding
  2. Hang the samples on turntable according to the requirement of test, and number them.
  3. Add the prepared etching solution to the solution tank; about 70 liters solution is needed.
  4. Add an appropriate amount of distilled water into the humidification water tank and condensate pipe, and connect the condensate drain to the wastewater barrel.
  5. During operation, please do not open the door or put your hand into the test chamber unless it is quite necessary.

 

II. Parameter setting and experiment

  1. Power on, and set parameters according to screen prompts
  2. Press the "Run the Program" and "Run the value" to set temperature, humidity, dial speed and other related parameters.
  3. After the parameter setting, press "Run" button to start the experiment. During the experiment, monitor the liquid consumption in solution tank, humidifying water tank and condensate water. Replenish in time.
  4. When the experiment is conducted to a certain period or needs to increase the solution, open the chamber door, take out the sample quickly and immediately shut the door. The door open time should be less than 2 minutes.
  5. Power off after experiment. Remove all samples and empty the solution tank. Wash the tank for two times.

 

 

 