High-vacuum carbon coater
Name: High-vacuum carbon coater
Model: Cressington 208C high
Manufacturer: UK Cressington
Purchasing date: December 23, 2009
Main Specifications:
(1) Thickness measuring range: 0 ~ 999.9nm; Accuracy: 0.1nm
(2) Density range: 0.5 ~ 30 gm / cm3;
(3) Immersion bell size: 150mm diameter
(4) Height: 165 ~ 250mm
(5) Carbon type: 6.15mm carbide, stainless steel fixtures
(6) Sample stage: asynchronous planetary transmission stage
Common accessories: pachymeter
Main application area: sample preparation equipment
Main functions: Spraying carbon film to the sample surface of scanning electronic microscopy, electronic microprobe and other equipment
Person in charge of the equipment: Wu Hongyan
Ways of contact:
Tel: 024-83690365
Email:
Location: Room 107, RAL
Operation Rules
- Connect the power supply and turn on the host machine.
- After the green light flashes, wait for another 3~5 minutes to adjust the voltage (4v) and time (6s)
- Turn on the small switch on the back of the host, and adjust the rotating speed to 3~4 level and set voltage and time.
- Open and reset the pachymeter.
- Press ‘START’ to plate carbon
- Stop the machine and turn off the small switch behind the host before completing the test.
- Press ‘POWER’
- After powering off, open the upper-cover of the glass shield. Cover it back till the shield temperature cooling down to room temperature.